The Owandy-RX AC was the alternating-current version of the Owandy-RX intraoral generator, available as a wall-mounted unit (upper or lower bracket) or as a mobile unit on a cart. A sturdy, vibration-free device compatible with film, digital sensors and phosphor plate systems, it equipped thousands of offices before being withdrawn from the catalog in August 2026 in favor of the high-frequency DC version.
Step-by-step wall installation of the Owandy-RX generator: timer, bracket, pantograph arm and tube head. The video shows the DC version, whose mechanics, arms and tube head are identical to those of the Owandy-RX AC.
Its microprocessor timer drove up to two tube heads and automatically corrected the exposure time according to line fluctuations, for a constant dose and no retakes. The one-piece light-alloy tube head, oil-immersed, housed the tube at the rear of the casing for a source-to-skin distance 50% greater than conventional designs; the pantograph arm, adjustable in height and depth, came in three lengths.

Single-phase · 70 kV · 8 mA · exposure time 0.08 to 3.2 s in 17 steps · duty cycle 1:32.
Microprocessor, automatic line compensation · up to 2 tube heads · confirmed film / digital selection.
Toshiba DG-073 or CEI OX70 tube, 0.7 mm focal spot · 12 lb (5.5 kg) one-piece head, 300° rotation · increased source-to-skin distance.
Short 8 in (20 cm), long 12 in (31 cm), rectangular 44 × 35 mm (12 in) · bisecting and paralleling techniques.
Pantograph arm in 3 lengths (16, 31, 43 in) · high or low wall mounting · 110 lb mobile version · 43 lb wall-mounted set.
115 / 220 / 230 V, 50/60 Hz · class I type B · class IIb medical device · remote control on extendable cable > 6 ft.
User manual (09/2024 edition) in six languages, in PDF format.
For any other documentation, please contact U.S. technical support.
Frequently asked questions from Owandy-RX AC users.
Yes. The product has not been sold since August 2026 but it is still recent: parts, maintenance and exposure factor checks are provided by Owandy Inc. Contact U.S. technical support and quote the serial number.
The user manual (September 2024 edition) is available above in English, French, Spanish, Italian, German and Portuguese.
Yes: the generator is compatible with film, digital sensors (ONE, Opteo) and phosphor plates (Owandy-CR). The timer offers separate exposure values for film and digital.
With the Owandy-RX DC, the high-frequency version: same arms, same cones and same ergonomics, with variable high voltage from 60 to 70 kV, exposure times accurate to 1/100th of a second and an optional wireless trigger.


The Owandy-RX DC keeps the frame, arms and cones of the AC version and moves to high frequency: